Wavecom Instruments Fully Integrated Test n Print Unit User Manual
15
TnP Options Menu
Introduction
Here you can adjust the way the TnP behaves under certain tests and other functions. Some TnP options may
not available to some TnP versions.
To access the TnP options menu:
From either the Enter barcode screen (memory unit) or the main menu ‘A ‘ screen hold down the enter
button, and then press the F2 key
You should now be in the TnP options screen. The screen displays 1 option at a time.
You can select this option by pressing the enter button.
To select the next item press the F2 button.
Pressing the F3 button escapes out of the TnP options menu.
The TnP Memory model options list.
Change Insulation Voltage
Leakage Test Time
Audio Option
Data Results Screen Display Hold Time
Label Options
Barcode Options
Change Insulation Voltage - All TnP models
Some DUT contain MOV's (metal oxide varistor) or commonly known as spike protection devices. These limit the
mains voltage to about 280VAC. If you attempt to perform a 500VDC insulation test on one of these devices you
will get a failed result. To fix this problem select the 250V option.
Press the key to select the 250V option.
Press the key to select 500V option.
You can also set the unit back to 500V by switching the mains power off and on again. The default insulation
voltage is 500V.