Function
Range
Resolution
Accuracy
Resistance
600
0.1
(0.3% reading + 4 digits)
6k
0.001k
60k
0.01k
600k
0.1k
6M
0.001M
60M
0.01M
(0.5% reading + 20 digits)
Capacitance
60nF
0.01nF
(3.5% reading + 4 digits)
600nF
0.1nF
(5% reading + 5 digits)
Frequency
(electronic)
9.999Hz
0.001Hz
(0.1% reading + 1 digits)
99.99Hz
0.01Hz
999.9Hz
0.1Hz
9.999kHz
0.001kHz
99.99kHz
0.01kHz
999.9kHz
0.1kHz
9.999MHz
0.001MHz
40MHz
.01MHz
Sensitivity: 0.8V rms min. @ 20% to 80% duty cycle and <100kHz; 5Vrms min @
20% to 80% duty cycle and > 100kHz.
Frequency
(electrical)
10.00-400Hz
0.01Hz
(0.5% reading)
Sensitivity: 15Vrms
Duty Cycle
0.1 to 99.9%
0.1%
(1.2% reading + 2 digits)
Pulse width: 100µs - 100ms, Frequency: 5Hz to 150kHz
Temp
(type-K)
-50 to 1382 F
1 F
(3.0% reading + 5 C /9 F digits)
(probe accuracy not included)
-45 to 750 C
1 C
Note: Accuracy specifications consist of two elements:
(% reading) This is the accuracy of the measurement circuit.
(+ digits) This is the accuracy of the analog to digital converter.